Abstract
An (e, 2e) spectrometer has been constructed for detecting scattered and recoiling electrons from a thin solid film. The cross section for observing the two electrons in coincidence is proportional to the spectral momentum density of the target. In this spectrometer the energy of the incident electron beam is 25 keV and the beam current is approximately 40 μA. The energy resolution (FWHM) is ≤4 eV. The momentum resolution (FWHM) can be varied from 0.2 to 1.0 Å−1. Preliminary coincidence data from an amorphous carbon film are presented.
Recommended Citation
A. L. Ritter, JR Dennison, and J. Dunn, "An (e,2e) Spectrometer for Investigating the Spectral Momentum Density of Thin Films," Rev. Sci. Instrum. 55(8), 1280-1286 (1984).