7-2-2008

Abstract

This paper details the design and validation of a Multiple Path OP-FTIR system with elevation and radial scanning ability and demonstrates its capabilities to quantify and monitor gaseous ammonia emitted from agricultural facilities. The OP-FTIR system has a 500 m range (1000 m full path length) and allows 360° radial scan and 45° scan in elevation. To study large scale sources, two or more similar systems may be needed. For comparison purposes, we ran two similar but not identical OP-FTIR systems side-by-side in a controlled lab environment and in a series of field environments. We determined that in a controlled environment, the two systems can attain an NH3 agreement of 1- 3% at concentrations above 500 ppb. Due to the short path length (~10 m) in the lab, 500 ppb was the detection limit of the two systems. Path lengths in a field are much longer, allowing a lower detection limit. Average agreement in the field was 1-6%. This difference in agreement from the laboratory is likely due to the non-homogeneous distribution of the pollutant.