Files
Download Full Text (546 KB)
Description
The purpose of this project is to design and develop a testing-apparatus and test-method to measure the maximum allowable leakage limit (MALL) for films used by Fresenius Medical Care (FMC). The MALL is the greatest leak size that does not pose a risk to the product. This is the first step in developing container closure integrity testing for sterile containers. This test is designed using guidance from USP 1207, a document detailing integrity assurance of packages holding sterile medical products, and 21 CFR 211. This test is focusing on microbial ingress rather than physical or chemical leakage. The test-method design is based on an aerosol microbial ingress test published in the PDA Journal of Pharmaceutical Science and Technology [1]. A few of the key variables considered are film, microbes, channel diameters, and pressure.
Publication Date
12-6-2022
City
Logan, UT
Keywords
dialysis films, integrity assurance, leakage limit, microbial ingress
Disciplines
Engineering
Recommended Citation
Phillips, Lexi, "Developing a Test Method to Determine the Maximum Allowable Leakage Limit of Microbial Ingress for Dialysis Films" (2022). Fall Student Research Symposium 2022. 61.
https://digitalcommons.usu.edu/fsrs2022/61