Files

Download

Download Full Text (546 KB)

Description

The purpose of this project is to design and develop a testing-apparatus and test-method to measure the maximum allowable leakage limit (MALL) for films used by Fresenius Medical Care (FMC). The MALL is the greatest leak size that does not pose a risk to the product. This is the first step in developing container closure integrity testing for sterile containers. This test is designed using guidance from USP 1207, a document detailing integrity assurance of packages holding sterile medical products, and 21 CFR 211. This test is focusing on microbial ingress rather than physical or chemical leakage. The test-method design is based on an aerosol microbial ingress test published in the PDA Journal of Pharmaceutical Science and Technology [1]. A few of the key variables considered are film, microbes, channel diameters, and pressure.

Publication Date

12-6-2022

City

Logan, UT

Keywords

dialysis films, integrity assurance, leakage limit, microbial ingress

Disciplines

Engineering

Developing a Test Method to Determine the Maximum Allowable Leakage Limit of Microbial Ingress for Dialysis Films

Included in

Engineering Commons

Share